卷曲霉素在罗氏培养基中的稳定性研究

江勇 钱明 王威 陈涛 李海成 陈燕梅 卓文基 郭卉欣 周琳 钟球   

  1. 510630  广州,广东省结核病控制中心(江勇、钱明、陈涛、李海成、陈燕梅、卓文基、郭卉欣、周琳、钟球);广东省佛山市第四人民医院检验科(王威)
  • 收稿日期:2013-06-06 出版日期:2013-08-10 发布日期:2013-08-04
  • 通信作者: 周琳;钟球 E-mail:gdtb_bg@vip.163.com;zhongqiu@vip.163.com
  • 基金资助:

    “十二五”国家科技重大专项(2013ZX10003001)

The research on capreomycin stability in Roche medium

JIANG Yong,QIAN Ming,WANG Wei,CHEN Tao,LI Hai-cheng,CHEN Yan-mei,ZHUO Wen-ji,GUO Hui-xin,ZHOU Lin,ZHONG Qiu   

  1. Center for Tuberculosis Control and Prevention of Guangdong Province, Guangzhou 510630, China
  • Received:2013-06-06 Online:2013-08-10 Published:2013-08-04
  • Contact: ZHOU Lin;ZHONG Qiu E-mail:gdtb_bg@vip.163.com;zhongqiu@vip.163.com

摘要: 目的  检测卷曲霉素(Cm)在罗氏培养基中的稳定性,了解Cm在罗氏培养基中的降解速率;检测Cm降解对药物敏感性试验结果的影响。 方法  用高效液相色谱仪分别检测4 ℃和37 ℃ 2种保存条件下,Cm在罗氏培养基中的稳定性;使用不同保存条件下的罗氏培养基进行药物敏感性试验,以验证Cm降解对敏感性试验结果的影响。 结果  在4 ℃保存时,罗氏培养基中的Cm含量与时间呈反比,保存10 d时,罗氏培养基中的Cm降解2.04%,保存40 d时,Cm降解21.5%,保存80 d时,Cm降解41.7%,保存到200 d时,检测不到Cm的存在;在37 ℃保存10 d时,罗氏培养基中的Cm降解46.8%;在37 ℃保存30 d时,无法在罗氏培养基中检测到Cm的存在。在4 ℃存储时间为10~160 d的Cm罗氏培养基进行结核分枝杆菌药物敏感性试验(DST)时,MIC在0.8~1.0 μg/ml菌株,在4 ℃存储时间为40 d的Cm罗氏培养基上生长6周后出现16个菌落。MIC在1.0~1.5 μg/ml菌株,在4 ℃存储时间为40 d的Cm罗氏培养基上生长6周后出现阳性结果(+)。 结论 Cm罗氏培养基在37 ℃保存时不稳定,在4 ℃保存时相对稳定,但存储时间不应超过40 d。

关键词: 分枝杆菌, 结核, 培养基, 微生物敏感性试验, 卷曲霉素硫酸盐, 色谱法, 高压液相

Abstract: Objective To detect the stability of capreomycin (Cm) in the Roche medium, and find out the degradation rate of Cm in Roche medium and to detect the influence of Cm degradation on drug sensitive test. Methods The high performance liquid chromatograph(HPLC) test was used to detect the stability of Cm under preservation condition of 4 ℃ and 37 ℃ in Roche medium; Cm under different preservation condition was used to conduct drug sensitive test (DST) to detect the influence of Cm degradation on drug sensitive test. Results Preserved in 4 ℃, the concentration of Cm in Roche medium is inversely proportional to time, the degradation of Cm in Roche medium is close to 2.04% after 10 days, the degradation is 21.5% after 40 days, the degradation is 41.7% after 80 days, and Cm was not detected after 200 days. Preserved in 37 ℃, the degradation of Cm in Roche medium is close to 46.8% after 10 days, and Cm was not detected after 30 days. In the Cm Roche medium preserved at 4 ℃ for 10-160 days, DST results showed that in the strain which minimal inhibitory concentration (MIC) is between 0.1 to 0.2 μg/ml, 16 colonies were observed after 6 weeks in the Roche medium preserved in 4 ℃ for 40 days. In the strain which MIC is between 1.0 to 1.5 μg/ml, positive result (+) was reported after 6 weeks in the Roche medium preserved in 4 ℃ for 40 days. Conclusion Cm Roche medium is unstable when stored at 37 ℃, it should be preserved at 4 ℃ and should be used within 40 days.

Key words: Mycobacterium tuberculosis, Culture media, Microbial sensitivity tests, Capreomycin sulfate, Chromatography, high pressure liquid